Reverse Cycle Waterbath
 

reverse cycle waterbath

TECHNICAL SPECIFICATIONS

 

SB4VCH

SB4HCH

SB8CH

SB20CH

Temperature range  

ºC

0 - 65

 0 - 65

0 - 75

0 - 75

Temperature resolution  

±ºC

0.1

0.1

0.1

0.1

Ambient temperature  

ºC

32(10 - 43)

32(10 - 43)

32(10 - 43)

32(10 - 43)

Control type    

NTC probe

NTC probe

NTC probe

NTC probe

Heating capacity
 
0 - 27ºC
27 - 55ºC

W
W

 
140

 
140

440
300

 
600

Cooling capacity
 
 
55 - 27ºC
27 - 10ºC
10 - 0ºC

W
W
W

  
118
  

 
118
 

430
205
105

 
400
 

Power consumption 5ºC evaporating T

W

71

71

210

470

Pump:
 
Power consumption
max. pressure
max. flow rate
 

mbar
l/min
W

150
6
10

150
6
10

225
12
25

225
12
25

Bath:
 
Volume
Opening (WxLxH)

l
mm

4
230x130x150

4
230x130x150

8
280x230x150

20
490x280x150

Overall Dimensions W x L x H

mm

310x255x350

310x450x200

300x400x400

580x400x400

Total wattage (240V)  

W

100

100

250

505

CHILLER ONLY MODELS

 

SB4V

SB4H

SB8

SB20

Shaka Scientific reserves the right to alter the specifications without notice in the normal course of research and development.

reverse cycle bath graphs

Heating & cooling capacities (SB8CH)                    Heating & cooling curves (SB8CH)